[table id=4 /]

Prostate

Indications

  • Prostate Cancer
  • Infection
  • Elevated PSA
  • Lesion

Preparation/Positioning

Positioning: Supine Head First

Centering: Center on prostate

Coil: Body Surface Coil or Endo rectal coil

*Earplugs Mandatory*

Procedure: Make patient comfortable.

Dietary Prep: Patient may need to prep prior to exam to reduce gas and stool in rectum.

Sagittal T2 FSE

FOV

180mm

Image Matrix

256×256

Slice Thickness

3mm

Slice Gap

0mm

Coronal T2 FSE

FOV

180mm

Image Matrix

256×256

Slice Thickness

3mm

Slice Gap

0mm

Axial T2 FSE

FOV

180mm

Image Matrix

320×320

Slice Thickness

3mm

Slice Gap

0mm

Axial T1 Fat Sat FSE

FOV

180mm

Image Matrix

256×256

Slice Thickness

3mm

Slice Gap

0mm

Axial Diffusion

FOV

230mm

Image Matrix

160×160

Slice Thickness

3mm

Slice Gap

0mm

Axial T1 Whole Pelvis

FOV

300mm

Image Matrix

256×256

Slice Thickness

5mm

Slice Gap

0mm

Optional Coronal STIR Whole Pelvis

(If patient has history of prostate cancer)

FOV

300mm

Image Matrix

256×256

Slice Thickness

5mm

Slice Gap

0mm

If Ordered With Contrast, Add the Following

Axial Dynamic

FOV

180mm

Image Matrix

192×192

Slice Thickness

3mm

Slice Gap

0mm

Axial T1 Fat Sat FSE

FOV

180mm

Image Matrix

320×320

Slice Thickness

3mm

Slice Gap

0mm

Axial T1 Fat Sat FSE

(Whole Pelvis)

FOV

300mm

Image Matrix

320×320

Slice Thickness

5mm

Slice Gap

0mm

Plotting:

Option 1: Orthagonal

Option 2: Angle perpendicular to prostatic urethra

Considerations:

  • Set phase A-P and compensate for aliasing.
  • Saturation band inferior and superior to our FOV can help reduce artifacts.

Plotting:

Option 1: Orthagonal

Option 2: Angle parallel to prostatic urethra

Considerations:

  • Set phase H-F and compensate for aliasing.
  • Open RectFOV above 100% to increase coverage.
  • Saturation band inferior and superior to our FOV can help reduce artifacts.

Plotting:

Orthogonal in sagital plane

Considerations:

  • Set phase H-F and compensate for aliasing.
  • Open RectFOV above 100% to increase coverage.
  • Saturation band inferior and superior to our FOV can help reduce artifacts.